Data Memory |
32 MB |
Storage Capacity |
> 1.04 M arrays |
Communication Type |
USB, RS-232 |
Communication Speed |
115.2 kBits/second |
Communication Parameters |
8,N,1 (data bits, parity, stop bits) |
Communication Protocol |
Modbus RTU |
Scan Interval |
10-1440 minutes |
USB Driver (not applicable to RS-232 version) |
FTDI |
Power Supply |
Battery pack, or 9‑24V DC external |
Operating Time |
Please contact geokon |
Operating Temperature |
−40 °C to +85 °C |
(L×W×H) |
160 × 260 × 91 mm (Cellular Gateway + Mesh Supervisor) |
table 11: Supervisor Specifications
Data Memory |
32 MB |
Storage Capacity |
Varies by model |
Trueness |
0.082 Hz |
Frequency Precision |
±0.146 Hz (99% CI) |
Frequency Resolution |
±0.002 Hz |
Thermistor Accuracy |
External thermistor: ±0.5 °C |
Thermistor Resolution |
External thermistor: 0.032 °C |
Scan Interval |
10-1440 minutes |
Power Supply |
D cell, alkaline or lithium (2x), or 12V external |
Operating Temperature |
−40 °C to +85 °C (range varies by power source) |
VW Excitation Peak Current |
25 mA (max) |
VW Frequency Range |
400 Hz − 5000 Hz |
Sweep/Read Duration per Channel |
< 500 mS |
Battery Life |
Refer to Section 4.2 |
Dimensions (L×W×H) |
120 × 122 × 91 mm (single-channel node) |
Vibrating wire (VW) channel |
Gas discharge tube |
Thermistor (TH) channel
|
Gas discharge tube |
table 12: Node Specifications
Network |
8901-NA |
8903-EU |
Topology |
Star/Mesh/Cluster Tree (Automatic) |
Star/Mesh/Cluster Tree (Automatic) |
Radio Technology |
FHSS |
FHSS |
Radio Frequency, ISM Band |
902 − 928 MHz |
863 − 870 MHz |
Channels |
12 |
12 |
Range1 |
Up to 60 km (15 km x 4 hops) |
Up to 22 km (5.5 km x 4 hops) |
Transmit Power |
1 W |
20 mW |
Receiver Sensitivity |
−106 dBm |
−106 dBm |
Antenna (Half‑Wave Dipole) |
2.1 dBi |
1.6 dBi |
1Line-of-sight, maximum 4 hops |
table 13: Network Specifications
Precision |
±26.9 [0.0075] |
Arcseconds [Degrees] |
Linearity |
@±15° FS: ±115.5 [0.0321], |
Arcseconds [Degrees] |
Temperature‑Dependent Uncertainty |
68.8 [0.019 |
Arcseconds [Degrees]/°C |
Angle Resolution |
0.9 [0.00025] |
Arcseconds [Degrees] |
Tilt Range |
±90 |
Degrees |
Axes |
2 |
|
table 14: Tiltmeter Specifications